Comet H3021P compressed air temperature and humidity regulator

Previous 6 / 145 Next

Comet H3021P compressed air temperature and humidity regulator Transmitters and Regulators Comet
Quantity
Add Quotation


Download Files

Cometsystem_H3021P.pdf (96.49 KB)
code: H3021P
Humidistat, thermostat with two relay outputs. T+RH probe from hard anodized duralumin for compressed air up to 25 bars. Outdoor, indoor use.
 

Humidity, temperature probe on a cable. Measured values are also converted to other humidity interpretation: dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy.

The device is supplied with T+RH probe with 1m cable. Cable lengths 2m or 4m available optionally.

 Included in delivery: 

  • Traceable calibration certificate
  • Quick start manual
  • Free program TSensor for configuring of the transmitter

Technical parameters Value
Output None
Measured Value Temperature + Relative Humidity
Construction Type Compressed Air Humidity up to 2.5MPa
Design Industrial
Temperature Measuring Range -30 to 105 °C
Relay Output Yes
Two-State Input No
Lcd Display Yes
PoE No
Maximum switching voltage, current, power of relay output 50V, 2A, 60VA
Audible alarm from built-in beeper - switchable
Relative humidity range 0 to 100%
Accuracy of relative humidity measurement ±2.5% relative humidity from 5 to 95% at 23°C
Accuracy of temperature output ±0.4°C from -30 to +100°C
±0.4% from reading over +100°C
Available temperature units degrees Celsius, Fahrenheit
Dew point temperature range -60 to +80 °C
Accuracy of dew point temperature output (for more details see instruction manual) ±1.5°C for dew point temperature +10°C and higher at ambient temperature +25°C
Accuracy of dew point temperature output ±2.0°C for dew point temperature 0°C at ambient temperature +25°C
Accuracy of dew point temperature output ±3.0°C for dew point temperature -10°C at ambient temperature +25°C
Accuracy of dew point temperature output ±6.0°C for dew point temperature -20°C at ambient temperature +25°C
Accuracy and range of absolute humidity output ±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3
Accuracy and range of specific humidity output ±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg
Accuracy and range of mixing ratio output ±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg
Accuracy and range of specific enthalpy output ±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg
Temperature operating range -30 to +80°C
Temperature operating range of LCD display readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
Temperature compensation of the humidity sensor all temperature range
Filtering ability of sensor cover 0.025mm - filter with stainless steel mesh
IP protection IP65 electronics, IP40 sensors
Power 9-30Vdc, power consumption approximately 1W
Mechanical connection of probe G1/2 with O-Ring
Length of the probe cable 1m, 2m or 4m optionally
Dimensions 136 x 159 x 45 mm (W x H x D), probe length 150 mm, probe diameter 18mm
Weight approximately 450g

Optional accessories

Description
AC/DC adapter 12V/450mA stabilized
cable for transmitter adjustment via USB port
Wall holder for external probes DIGIL
plastic sensor protection with stainless filter (black)
PTFE sensor protection, white
Flow chamber for compressed air measurement up to 25 bars
Tool for easy connection to WAGO terminals 233 for serial output transmitters
Vessel for adjustment and calibration of humidity
Humidity Standard 10%RH, set 5pcs
Humidity Standard 80%RH, set 5pcs